Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Metrics
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HWeakness enumeration
Affected software
No affected versions have been published.
Timeline
- Published
- Mar 25, 2019
- Last modified
- Jun 17, 2026
- NVD status
- Modified