Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around: Only authorized people can physically access the analyzer.
Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Metrics
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HWeakness enumeration
Affected software
| Vendor | Product | Versions |
|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | Windows 7 Operating system; Windows XP Operating system |
| Radiometer Medical Aps | AQT90 FLEX Analyzers | Windows 7 Operating system; Windows XP Operating system |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX analyzers | Windows 7 Operating system; Windows XP Operating system |