Description
A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration.
Metrics
CVSS 3.1 7.5/10
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:NWeakness enumeration
Affected software
| Vendor | Product | Versions |
|---|---|---|
| n/a | n/a | n/a |
References
- https://www.bbraunusa.com/en.htmBroken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/Exploit, Third Party Advisory
- https://www.bbraunusa.com/en.htmBroken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/Exploit, Third Party Advisory
Timeline
- Published
- Aug 25, 2021
- Last modified
- Jun 17, 2026
- NVD status
- Modified