In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a hard-coded physician PIN in the physician menu of the insulin pump allows attackers with physical access to change insulin therapy settings.
Metrics
CVSS 3.1 6.8/10
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HWeakness enumeration
Affected software
No affected versions have been published.
Timeline
- Published
- Jan 19, 2021
- Last modified
- Jun 17, 2026
- NVD status
- Modified