Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Metrics
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:LWeakness enumeration
Affected software
Source: the vendor's advisory (CNA). NVD analysis pending.
No affected versions have been published.
Timeline
- Published
- Jun 2, 2026
- Last modified
- Jul 22, 2026
- NVD status
- Awaiting Analysis