Description
Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Metrics
CVSS 3.1 5.3/10
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:NWeakness enumeration
Affected software
| Vendor | Product | Versions |
|---|---|---|
| Medtronic | MMT- 508 - MiniMed pump | All versions |
| Medtronic | MMT – 511 pump Paradigm | All versions |
| Medtronic | MMT – 512 / MMT – 712 Paradigm x12 | All versions |
| Medtronic | MMT – 515 / MMT – 715 Paradigm x15 | All versions |
| Medtronic | MMT – 522 / MMT – 722 Paradigm REAL-TIME | All versions |
| Medtronic | MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME | All versions |
| Medtronic | MMT – 523 / MMT – 723 Paradigm Revel | All versions |
| Medtronic | MMT – 523(K) / MMT – 723(K) Paradigm | All versions |
| Medtronic | MMT – 554 / MMT – 754 MiniMed Veo | All versions |
| Medtronic | MMT – 551 / MMT – 751 MiniMed 530G | All versions |
References
- http://www.securityfocus.com/bid/105044Third Party Advisory, VDB Entry
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/105044Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02Third Party Advisory, US Government Resource
Timeline
- Published
- Aug 13, 2018
- Last modified
- Jun 17, 2026
- NVD status
- Modified