Description
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.
Metrics
CVSS 3.0 4.8/10
CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:LWeakness enumeration
Affected software
| Vendor | Product | Versions |
|---|---|---|
| n/a | Ethicon Endo-Surgery Generator G11 | Ethicon Endo-Surgery Generator G11 |
References
- http://www.securityfocus.com/bid/101978Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-332-01Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/101978Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-332-01Third Party Advisory, US Government Resource
Timeline
- Published
- Dec 5, 2017
- Last modified
- Jun 17, 2026
- NVD status
- Modified