Description
Johnson & Johnson Animas OneTouch Ping devices do not properly generate random numbers, which makes it easier for remote attackers to spoof meters by sniffing the network and then engaging in an authentication handshake.
Metrics
CVSS 3.0 7.5/10
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:NWeakness enumeration
Affected software
| Vendor | Product | Versions |
|---|---|---|
| n/a | n/a | n/a |
References
- http://www.kb.cert.org/vuls/id/884840Third Party Advisory, US Government Resource
- http://www.kb.cert.org/vuls/id/BLUU-A9SQRSThird Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/93351
- https://community.rapid7.com/community/infosec/blog/2016/10/04/r7-2016-07-multiple-vulnerabilities-in-animas-onetouch-ping-insulin-pumpMitigation, Technical Description, Third Party Advisory
- https://ics-cert.us-cert.gov/advisories/ICSMA-16-279-01
- http://www.kb.cert.org/vuls/id/884840Third Party Advisory, US Government Resource
- http://www.kb.cert.org/vuls/id/BLUU-A9SQRSThird Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/93351
- https://community.rapid7.com/community/infosec/blog/2016/10/04/r7-2016-07-multiple-vulnerabilities-in-animas-onetouch-ping-insulin-pumpMitigation, Technical Description, Third Party Advisory
- https://ics-cert.us-cert.gov/advisories/ICSMA-16-279-01
Timeline
- Published
- Oct 5, 2016
- Last modified
- Jun 17, 2026
- NVD status
- Modified