Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Metrics
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:NWeakness enumeration
Affected software
No affected versions have been published.
Timeline
- Published
- Mar 25, 2019
- Last modified
- Jun 17, 2026
- NVD status
- Modified