Description
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Metrics
CVSS 2.0 9.3/10
AV:N/AC:M/Au:N/C:C/I:C/A:CWeakness enumeration
Affected software
| Vendor | Product | Versions |
|---|---|---|
| Hospira | LifeCare PCA Infusion System | <= 5.0 |
References
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmThird Party Advisory, US Government Resource
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmThird Party Advisory, US Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01Third Party Advisory, US Government Resource
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
Timeline
- Published
- Jul 6, 2015
- Last modified
- Jun 17, 2026
- NVD status
- Modified