Description
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Metrics
CVSS 2.0 9.3/10
AV:N/AC:M/Au:N/C:C/I:C/A:CWeakness enumeration
Affected software
No affected versions have been published.
Timeline
- Published
- Jul 6, 2015
- Last modified
- Jun 17, 2026
- NVD status
- Modified