Description
In the default configuration, the HL7 listener accepts ADT^A40 patient-merge messages without authenticating the sending application. A remote sender can reassign studies and related workflow records from one patient identity to another. The archive's unmerge does not reverse this.
Metrics
CVSS 3.1 7.5/10
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:NWeakness enumeration
Affected software
No affected versions have been published.
Timeline
- Published
- Oct 16, 2026
- Last modified
- Oct 16, 2026
- NVD status
- Deferred