Description
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Metrics
CVSS 3.1 5.7/10
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:NWeakness enumeration
Affected software
| Vendor | Product | Versions |
|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All |
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01Mitigation, Third Party Advisory, US Government Resource
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01Mitigation, Third Party Advisory, US Government Resource
Timeline
- Published
- Sep 13, 2022
- Last modified
- Jun 17, 2026
- NVD status
- Modified